News
November 11, 2013
How to Compute the Mean Particle Diameter from a LISST Volume Distribution
[Sequoia, March 26, 2010]
In the table below, the computation of the mean particle size from a LISST volume distribution is outlined.Column A is the size class...
Read more...
October 11, 2013
The acceptance angle of the LISST-100X / LISST-STX transmission measurement differs between instruments:
IN AIR, the LISST-100X type C has an acceptance angle of 0.0177 degrees while a type B has an acceptance angle of 0.0353 degrees.
The discontinued...
Read more...
September 23, 2013
For the LISST-100X and LISST-Portable particle sizers, the user has an option to choose between two different shape models for inverting the scattering pattern into a size distribution. The two models are called Spherical and Randomly shaped particles.
The...
Read more...
August 30, 2013
Yes. The files called factory_zsc_xxxx.asc, ringarea_xxxx.asc (where xxxx is the serial number) are instrument specific.
Also the files InstrumentData.txt and LISST.ini contains instrument specific information, but these two files can contain information...
Read more...
January 9, 2013
Turbidity shows no response to large particles, gets concentration wrong
[Sequoia, January 09, 2013]
The use of turbidity sensors is widely prevalent in the sediment community. The common practice is to calibrate the turbidity meter...
Read more...