LISST-SL velocity thresholds

October 11, 2013

Is there a low velocity threshold for the LISST-SL? Does it need a minimum flow speed in order to sample? What is that? Absolutely! The internal pump in the SL will adjust the intake velocity to match the ambient flow speed in order to make the sample...

Read more...

What is the acceptance angle of the LISST instruments?

October 11, 2013

The acceptance angle of the LISST-100X / LISST-STX transmission measurement differs between instruments: IN AIR, the LISST-100X type C has an acceptance angle of 0.0177 degrees while a type B has an acceptance angle of 0.0353 degrees. The discontinued...